Applications & Materials:
TRACER pXRF analyzers serve a wide range of applications and material needs from everyday point-and-shoot testing to the ever-changing complexities of advanced applications and research. Please contact our scientists to discuss your application.
Whether it is in art conservation and archaeology, for teaching, in food safety and agriculture applications, for geochemistry or advanced materials, TRACERs can thoroughly and accurately analyze the elemental composition and identify standard or complex materials.
TRACERs can analyze multiple sample types, including liquids, slurries, powders, soils, sediment, sludge, cellulose, polymers, paper, solids, metals and alloys.
TRACER 5i:
- 8 um Be Window
- Detects Na to U
- Analyzes Mg to U
- Ideal for art, archeology and advanced materials applications
TRACER 5g:
- 1 um Graphene Window
- Detects F to U (with He atmosphere)
- Analyses Na to U
- Ideal for geosciences applications
TRACER 5 series Technical Details
TRACER 5i Technical Details
Detector window: 8 μm Beryllium
Detector: Proprietary 20mm2 silicon drift detector with < 140 eV @ 250,000 cps Mn Kα; resolution for optimum light element analysis.
Analysis range: Detects from Na to U; analyzes from Mg – U
Size: 27.3 cm x 9.4 cm x 29.5 cm (10.75 in x 3.7 in x 11.6 in) L x W x H
Weight: 1.9 kg (4.1 lbs) with battery; 1.6 kg (3.6 lbs) without battery
Power: Li-Ion battery and charger; AC adapter
Excitation source: Rhodium (Rh) thin window X-ray tube; X-ray generator 6-50 kV with 4.5-195 uA, maximum 4 Watt output. Operator adjustable current and voltage for optimum excitation
Collimation: User changeable collimator; 3 mm and 8 mm collimators supplied
Filters: Operator controlled, motorized 5 position primary beam filter wheel. Manual insertion filter/secondary target slot for factory or user made filters.
Geometry: Features patented SharpBeam™ beam path for best performance at low power.
Beam path: Capable of selectable beam path of vacuum, helium or air to detect elements as light as Fluorine (F) to as heavy as uranium (U).
Internal camera: Internal VGA CMOS camera with ability to store up to 5 photos per assay.
Interactive touchscreen: High performance and contrast daylight visible TFT LCD 9.4 cm (3.7”) color touchscreen display.
Convenience features: TrueTouch trigger switch, relaxed ambidextrous hand grip strap and EasyAccess rail mount for accessories.
Control software: Full control OS on analyzer and on PC software.
Analysis software: Control of all excitation parameters and live spectra displayed on analyzer. Live spectra and qualitative, semi-quantitative and quantitative analysis on Windows® 7 or 10 PC.
Data storage & data transfer: Internal and direct storage on USB drive. Data transfer to PC via USB or Wi-Fi. Bluetooth connectivity for accessories.
System safety: Password protection; Sample proximity sensor; Low count rate (backscatter) shutoff
Languages: Chinese, Chinese simplified, Dutch, English, French, French Canadian, German, Indonesian, Italian, Japanese, Korean, Polish, PortugueseBR, Russian, SpanishMEX, SpanishSPN, Thai, Turkish
Operating environment: -10°C to +50°C (+14°F to 122°F); IP54 dust and splash resistant; Altitude: 2,500 m (8,200 ft)
Certification: CE, FCC part 15
TRACER 5g Technical Details
Detector window: 1 μm Graphene
Detector: Proprietary 20mm2 silicon drift detector with < 140 eV @ 250,000 cps Mn Kα; resolution for optimum light element analysis.
Analysis range: Detects from F to U (with He atmosphere); analyzes from Na – U
Size: 27.3 cm x 9.4 cm x 29.5 cm (10.75 in x 3.7 in x 11.6 in) L x W x H
Weight: 1.9 kg (4.1 lbs) with battery; 1.6 kg (3.6 lbs) without battery
Power: Li-Ion battery and charger; AC adapter
Excitation source: Rhodium (Rh) thin window X-ray tube; X-ray generator 6-50 kV with 4.5-195 uA, maximum 4 Watt output. Operator adjustable current and voltage for optimum excitation
Collimation: User changeable collimator; 3 mm and 8 mm collimators supplied
Filters: Operator controlled, motorized 5 position primary beam filter wheel. Manual insertion filter/secondary target slot for factory or user made filters.
Geometry: Features patented SharpBeam™ beam path for best performance at low power.
Beam path: Capable of selectable beam path of vacuum, helium or air to detect elements as light as Fluorine (F) to as heavy as uranium (U).
Internal camera: Internal VGA CMOS camera with ability to store up to 5 photos per assay.
Interactive touchscreen: High performance and contrast daylight visible TFT LCD 9.4 cm (3.7”) color touchscreen display.
Convenience features: TrueTouch trigger switch, relaxed ambidextrous hand grip strap and EasyAccess rail mount for accessories.
Control software: Full control OS on analyzer and on PC software.
Analysis software: Control of all excitation parameters and live spectra displayed on analyzer. Live spectra and qualitative, semi-quantitative and quantitative analysis on Windows® 7 or 10 PC.
Data storage & data transfer: Internal and direct storage on USB drive. Data transfer to PC via USB or Wi-Fi. Bluetooth connectivity for accessories.
System safety: Password protection; Sample proximity sensor; Low count rate (backscatter) shutoff
Languages: Chinese, Chinese simplified, Dutch, English, French, French Canadian, German, Indonesian, Italian, Japanese, Korean, Polish, PortugueseBR, Russian, SpanishMEX, SpanishSPN, Thai, Turkish
Operating environment: -10°C to +50°C (+14°F to 122°F); IP54 dust and splash resistant; Altitude: 2,500 m (8,200 ft)
Certification: CE, FCC part 15